SpotOn Compact

High accuracy compact beam positioning system. No need for manifold box. Built-in electronics within sensor head.Measure both position and power, available with dual axis Lateral effect detector (down to 0.1um resolution @ 50Hz digital filter, +/-6um absolute accuracy), or with 4-Quadrant detector (accuracy down to +/-0.025% of beam diameter).

17 People watching this product now!

Description

SpotOn Compact

High accuracy compact beam positioning system. No need for manifold box. 

Built-in electronics within sensor head.

Measure both position and power, available with dual axis Lateral effect detector (down to 0.1um resolution @ 50Hz digital filter, +/-6um absolute accuracy), or with 4-Quadrant detector (accuracy down to +/-0.025% of beam diameter).

SpotOn-Compact-pic

Specifications

This product is designed to measure optical beam position of CW BEAMS ONLY

Detector L – Lateral Effect L04 Q – Four Quadrant
Size  9 x 9 mm  4 x 4 mm  10 x 10 mm 
Type  Dual axis Si, No glass cover  Dual axis Si, No glass cover  4x Si sectors seperated by 30 μm gap, No glass cover 
Usable beam size  ⌀50μm to ⌀9 mm  ⌀50μm to ⌀4 mm  ⌀50μm to ⌀5 mm 
Poisition measurement range  ⌀9 mm  ⌀4 mm  One beam radius up to ±2.5 mm 
Position resolution  Down to 0.1 μm at 50 Hz digital filter plus averaging  Down to 0.1 μm at 50 Hz digital filter plus averaging  Down to 0.25% of beam size 
Position accuracy  ±12.5 μm over 9 mm dia.  ±6 μm over 4 mm dia.  ±1μm or ±0.025% of beam size 
Update rate  Max 150 Hz at 500 Hz digital filter 
Spectral range  350 – 1100 nm 
Power range  1 μW to 10 mW with attenuating ND filters 
Power accuracy (*)  ±5% 

(*) – To maintain full calibration accuracy, attenuating optical filters may be necessary for operation with beams greater than 1 mW. Saturating “non-linear” effects depend on the beam size, type and wavelength, but caution should be exercised when using the 4Quad PSD above 3-6 mW or the Lateral Effect PSD above 1-3 mW.

Specification